Skip to main content

Table 3 Risk of error in accuracy results (limit 10% difference) calculated using linearity regression parameters (range 20 μg/mL to 200 μg/mL)

From: A risk-based approach to validation of ion chromatography methods using suppressed conductivity

Accuracy level

Percent difference in calculated concentration (μg/mL) using response factor at accuracy level shown

20

50

100

150

200

50% (25 μg/mL)

3.7

2.8

− 2.8

− 4.8

− 13.5

100% (50 μg/mL)

2.6

1.6

− 4.1

− 10.1

− 14.8

150% (75 μg/mL)

12.2

11.3

6.2

0.7

− 3.5