From: A risk-based approach to validation of ion chromatography methods using suppressed conductivity
Concentration (μg/mL) | Area | Statistics | RF | Statistics | ||
---|---|---|---|---|---|---|
35.06 | 0.654 |  |  | 1.866E−02 |  |  |
0.652 |  |  | 1.860E−02 |  |  | |
0.652 | N | 6 | 1.860E−02 | n | 6 | |
0.654 | SD | 0.001 | 1.866E−02 | SD | 2.80463E−05 | |
0.654 | Avg | 0.653 | 1.866E−02 | Avg | 1.86321E−02 | |
0.653 | %RSD | 0.15 | 1.863E−02 | %RSD | 0.15 | |
40.06 | 0.735 |  |  | 1.835E−02 |  |  |
0.739 |  |  | 1.845E−02 |  |  | |
0.747 | N | 6 | 1.865E−02 | n | 6 | |
0.740 | SD | 0.004 | 1.847E−02 | SD | 9.81621E−05 | |
0.742 | Avg | 0.741 | 1.852E−02 | Avg | 1.84871E−02 | |
0.741 | %RSD | 0.53 | 1.850E−02 | %RSD | 0.53 | |
50.08 | 0.927 |  |  | 1.851E−02 |  |  |
0.930 |  |  | 1.857E−02 |  |  | |
0.928 | n | 6 | 1.853E−02 | n | 6 | |
0.930 | SD | 0.001 | 1.857E−02 | SD | 3.09344E−05 | |
0.928 | Avg | 0.929 | 1.853E−02 | Avg | 1.85503E−02 | |
0.931 | %RSD | 0.15 | 1.859E−02 | %RSD | 0.17 | |
60.10 | 1.092 |  |  | 1.817E−02 |  |  |
1.091 |  |  | 1.815E−02 |  |  | |
1.104 | n | 6 | 1.837E−02 | n | 6 | |
1.082 | SD | 0.008 | 1.800E−02 | SD | 0.000133846 | |
1.084 | Avg | 1.092 | 1.804E−02 | Avg | 1.81626E−02 | |
1.096 | %RSD | 0.74 | 1.824E−02 | %RSD | 0.74 | |
65.10 | 1.172 |  |  | 1.800E−02 |  |  |
1.164 |  |  | 1.788E−02 |  |  | |
1.176 | n | 6 | 1.806E−02 | n | 6 | |
1.171 | SD | 0.004 | 1.799E−02 | SD | 5.96871E−05 | |
1.170 | Avg | 1.171 | 1.797E−02 | Avg | 1.79789E−02 | |
1.170 | %RSD | 0.33 | 1.797E−02 | %RSD | 0.33 |