Skip to main content

Table 5 Method precision results

From: A risk-based approach to validation of ion chromatography methods using suppressed conductivity

Rep

Theoretical concentration (μg/mL)

Area

Concentration found (μg/mL)

% Recovery

1

50.03

0.927

51.68

103.4

2

50.11

0.930

50.99

101.9

3

50.18

0.928

49.13

97.9

4

50.01

0.930

50.39

100.8

5

50.12

0.928

49.95

99.7

6

50.15

0.931

50.85

101.4

Avg

50.1

0.953

50.50

100.8

SD

0.067

0.015

0.890

0.902

%RSD

0.1

1.5

1.8

1.8