Table 6 Risk of an inaccurate result as a function of concentration calculated using the response factor for the 50 μg/ml working standard (limit 10% difference)

From: A risk-based approach to validation of ion chromatography methods using suppressed conductivity

Actual conc. (μg/mL)

Area

RF

Conc. found (μg/mL)

% Recovery

% Difference

35.06

0.653

0.01851

35.01

100.7

− 0.7

40.37

0.741

0.01851

40.37

99.1

0.9

50.19

0.929

0.01851

50.19

100.0

0.0

60.10

1.092

0.01851

60.07

98.1

1.9

65.10

1.171

0.01851

65.02

97.1

2.9